方晶晶,任小孟,徐新宏,江璐,王世锋,马骏,闫硕,方以群.高压环境下2种NDIR型CO2红外传感器的响应规律[J].装备环境工程,2022,19(7):129-136. FANG Jing-jing,REN Xiao-meng,XU Xin-hong,JIANG Lu,WANG Shi-feng,MA Jun,YAN Shuo,FANG Yi-qun.Response Law of Two Kinds of NDIR CO2 Sensor in Hyperbaric Environment[J].Equipment Environmental Engineering,2022,19(7):129-136.
高压环境下2种NDIR型CO2红外传感器的响应规律
Response Law of Two Kinds of NDIR CO2 Sensor in Hyperbaric Environment
  
DOI:10.7643/issn.1672-9242.2022.07.017
中文关键词:  高压  非分光型红外传感技术  二氧化碳  传感器  响应规律  信号漂移中图分类号:X703 文献标识码:A 文章编号:1672-9242(2022)07-0129-08
英文关键词:high pressure  non-dispersive infrared  carbon dioxide  sensor  response law  signal drift
基金项目:海军装备部基金(15A311)
作者单位
方晶晶 海军特色医学中心,上海 200433 
任小孟 海军特色医学中心,上海 200433 
徐新宏 海军特色医学中心,上海 200433 
江璐 海军特色医学中心,上海 200433 
王世锋 海军特色医学中心,上海 200433 
马骏 海军特色医学中心,上海 200433 
闫硕 海军特色医学中心,上海 200433 
方以群 海军特色医学中心,上海 200433 
AuthorInstitution
FANG Jing-jing Navy Medical Center, Shanghai 200433, China 
REN Xiao-meng Navy Medical Center, Shanghai 200433, China 
XU Xin-hong Navy Medical Center, Shanghai 200433, China 
JIANG Lu Navy Medical Center, Shanghai 200433, China 
WANG Shi-feng Navy Medical Center, Shanghai 200433, China 
MA Jun Navy Medical Center, Shanghai 200433, China 
YAN Shuo Navy Medical Center, Shanghai 200433, China 
FANG Yi-qun Navy Medical Center, Shanghai 200433, China 
摘要点击次数:
全文下载次数:
中文摘要:
      目的 研究压力变化环境对NDIR型CO2气体传感器的影响,获得不同压力范围内传感器的响应变化规律。方法 采用小型加压舱,在CO2分压不变的情况下,分别采用N2阶梯缓慢加压和冲击快速加压方式,研究英国C20和国产G4传感器在高压环境中稳定时间和漂移度等性能指标的波动规律,考察传感器对压力变化的响应规律。结果 在低于607 950 Pa的高气压稳压状态下,基本不会造成机械损伤,但信号会发生较大的漂移,国产G4分体式传感器出现信号震荡现象。CO2分体式传感器及整机的浓度输出值随着压力的升高,标准气样值与实测值的比值接近幂函数形式,分体式传感器的输出浓度值比整机的输出浓度值要大。结论 不同CO2浓度分压下的压力响应规律具有比较良好的一致性,冲击加压的漂移比阶梯加压显著,浓度真值可以通过测量总压和内置多次函数来修正。
英文摘要:
      The influence of pressure-changing environment on NDIR CO2 gas sensor was studied by using a small pressurized chamber, and the response variation law of the sensor in different pressure ranges was obtained thus. Under the condition of constant partial pressure of CO2, N2 static slow pressurization and impact fast pressurization were used respectively to study the fluctuation law of such performance indexes as stability time and drift of the two sensors, one is the C20 sensor made in Britain and the other is G4 sensor made in China The response law of the sensor to the changes of pressure and temperature was investigated, and the pressure compensation scheme under high pressure was proposed. When the pressure was lower than 60 meters, the mechanical damage did not occur but the signal drifted greatly. G4 sensor had a phenomenon of signal oscillation. The concentration output value of the two CO2 sensors and the whole machine increased with the pressure, and the ratio of the standard gas sample value to the measured value was close to the form of power function, besides, the output concentration value of the two sensors was larger than that of the whole machine. In conclusion, the pressure response law under different partial pressure of CO2 concentration shows good consistency. The signal drift of sensor under impact fast pressurization is more significant than that of N2 static slow pressurization. Moreover, the true value of concentration can be corrected by measuring total pressure and built-in multiple functions.
查看全文  查看/发表评论  下载PDF阅读器